How to Use Spine Curvature Correction



Step

1.

Select and switch between fusion pairs for review or calculation.

2.

By default, Adjust is activated.

Manually align the images using the move and rotate tools prior to calculation.

3.

Select Calculate to start the fusion.

Depending on the input images, you may be prompted to:

  • Select the image set to be corrected

  • Transfer the correction calculated for one image set to a network of further image sets

4.

By default, Spy Glass is activated.

Review the fusion by using Blending/Spy Glass.

5.

Select Scroll and scroll through the image slices using the arrow buttons .

6.

Review and verify the correction by toggling between Blending/Spy Glass.

Toggle Original to compare the curvature correction with the original image.

7.



Review and accept the fusion result if it is satisfactory.

8.



On accepting the fusion result, its status is displayed.

9.

Select Done when complete.

The corrected image set is saved for further processing.