How to Perform Software Revalidation

At least one DICOM patient should be available for this procedure.

Step

1.

Select a patient in Patient Browser.

2.

Select all relevant objects and contour a tumor object using SmartBrush.

3.

Start Adaptive Hybrid Surgery Analysis and wait until all fractionations are calculated.

4.

Switch between the different fractionation schemes.

5.

Check different organs at risk.

6.

Check the ACS, DVH and 3D views.

7.

Check that you can pan, zoom, scroll and perform windowing.

8.

Exit Adaptive Hybrid Surgery Analysis by selecting Done.

9.

Check the printout.

Revalidation is complete.